Aurius UV Vis Spectrophotometers Offer Optional GLP/GMP Validation Module
|
Performance Validation
Instrument performance may be validated using a comprehensive software
module available on all instruments and which may be released by
E-SEF.
Tests are provided for laboratories subject to regulatory control
or following good G.L.P. These include a range of Pharmacopoeia
tests (EP/BP/USP), a selection of which is shown here. |
| |
|
|
| |
|
|
Optical Bandwidth
The very accurate method for validating optical bandwidth uses
the narrow deuterium emission line at 656.1nm. The width, at half
height, of the measured band due to the emission line measures the
actual bandwidth. The band is drawn and the half width automatically
documented.
|
|
 |
| |
|
|
| |
|
|
Absorbance Validation
The Pharmacopoeia (EP/BP) test for absorbance accuracy is one of
the tests provided. The test measures the values of peaks and valleys
in the absorbance spectrum of a 60mg/L solution of Potassium Dichromate
in a 10mm pathlength cuvette. The results of such a test run are
shown here.
|
|
 |
| |
|
|
| |
|
|
Straylight
Straylight may be measured using the Pharmacopoeia (EP/BP) test
shown here. This test uses a 1.2% W/V solution of Potassium Chloride
in water, measured against water in the reference beam. An absorbance
in excess of 2A must be measured at a wavelength of 200nm.
The ASTMS tests at 220nm and 340nm are more widely used and are
provided for.
|
|
 |
| |
|
|
| |
|
|
Validation Using Filters
Absorbance and Wavelength validation are most conveniently carried
out using filter sets calibrated by Cecil Instruments on its reference
spectrophotometer against standards certified by the NPL. Full validation
programmes for these tests are provided.
|
|
 |