Aurius Specification |
| Monochromator |
 |
Littrow using 1200 L/mm holographic grating |
| Optical Bandwidth |
|
4nm |
| Wavelength Range |
|
190-1000nm; 325-1000nm visible versions |
| Wavelength Accuracy |
|
Better than ±1nm |
| Wavelength Reproducibility |
|
0.1nm |
| Self Test and Calibration |
|
Automatic at switch on |
| Straylight |
|
Typically less than 0.01% at 220nm and 340nm |
| Display Screen - LCD |
|
High resolution backlit with contrast control |
| Display Scrolling |
|
Provides 6 screen widths for viewing plots etc. |
| Autozero |
|
Automatic by press button |
| Photometric Ranges |
|
Digital display of -0.3-3A, 0-200%T, 0-9999C |
| Photometric Accuracy |
|
±0.005A or 1% whichever is greater |
| Photometric Noise |
|
Less than ±0.0002A (500nm) |
| Stability |
|
Better than 0.001A/hour at constant temperature |
| Multi-Wavelength Analysis |
|
2 and 3 wavelength methods. |
| Assay Program |
|
Enter multi-wavelength formula for automatic
computation and formatting of any assay |
| Curve Fitting and Editing |
|
Linear, quadratic or cubic; up to 30 standards |
| Cell and Wavelength Programming |
|
Up to 8 cells and 10 wavelengths combined |
| Method Storage |
|
Up to 30 methods stored in protected memory |
| Absorbance Ratioing |
|
260/280nm or two selected wavelengths |
| Peak Seek |
|
Automatic search programme, baseline corrected |
| Reaction Kinetics |
|
Tabulation of results, reaction plots, reprocessing |
| Batch Sampling |
|
100 samples, internal control, screen display |
| Real Time Clock |
|
Timed and dated reports |
| Computer/printer interfaces |
|
Bi-directional serial RS232 and parallel ports |
| Sample Temperature |
|
Displayed and set from keyboard |
| Size |
|
480 x 340 x 205mm |
| Weight |
|
16.8Kg Visible; 18.6Kg UV; + 0.9Kg with printer |
| Power Requirements |
|
110-250V, 50/60Hz, 170W/120W UV/Visible versions |
| |
|
|
SuperAurius Additional Specification |
| Optical Bandwidth |
|
1.8nm |
| Wavelength Scale Expansion |
|
Selectable by keyboard entry 1-100nm/cm |
| Scan Speed |
|
Selectable by keyboard entry 1-4000nm/m |
| Photometric Plotting Ranges |
|
Keyboard entry of limits; or automatic optimum |
| Baseline Flatness |
|
Better than ±0.002A (300nm/min) most of range |
| Overlayed Spectra |
|
Spectra and derivatives with or without offset |
| Spectral Reprocessing |
|
Spectra manipulated and replotted over any range |
| Spectral Storage |
|
Up to 100 spectra may be stored security code protected |
| Derivative Spectra |
|
1st and 2nd derivatives |
| Time Course Plotting |
|
Plots may be reprocessed and stored |