Skip to main content

HPLC Flow Meters

FlowCal 5000 verifies pump accuracy for IQ, OQ, and PQ with fast, repeatable measurements.

FlowCal 5000 digital HPLC flow meter
  • Verify pump accuracy for IQ
  • OQ
  • PQ with fast, repeatable measurements

There are no products listed under this category.

What digital HPLC flow meters do

Digital flow meters measure HPLC pump accuracy in fractions of milliliters per minute. Accurate flow delivers reproducible retention times and keeps regulated labs compliant.

Why it matters

HPLC qualification depends on precise pump flow. Flow meters replace slow volumetric or gravimetric checks with quick, optical readings that remove operator variability.

  • Supports FDA IQ/OQ/PQ documentation
  • Quarterly pump verification recommended by instrument makers
  • Built for milliliter-per-minute accuracy, not industrial flow ranges
Digital HPLC flow meter with accessories

Why labs choose the FlowCal 5000

Fast

Readings in seconds instead of 10-15 minutes per point with manual volumetric or gravimetric methods.

Accurate

Optical sensing removes meniscus reading error; linear across full flow range of 0.05-25 mL/min.

Regulatory ready

Traceable calibration and data output for audit trails and IQ/OQ/PQ packages.

Portable

Lightweight and compact; one flow meter services multiple systems or sites.

Metrologists who have used the FlowCal 5000 never want to go back to manual methods!

How the FlowCal 5000 works

FlowCal 5000 digital HPLC flow meter

Measurement sequence

  1. Solvent enters the inlet; first optical sensor starts the timer.
  2. Meniscus reaches the second sensor; timer stops, valve opens, and tube drains.
  3. Reading appears on the 4-digit LCD and can output to 6 digits over RS232 to PC or printer.
  4. Cycle repeats automatically for fast averaging.

FlowCal 5000 key specs

Compact, accurate, and ready for every HPLC system.

Performance

  • Accuracy: 1% of reading; Reproducibility 0.5%; Linear from 0.05-25 mL/min
  • Measurement volume: 140 μL; priming 250 μL
  • Reading update time:
    • 15 s @ 2 mL/min
    • 25 s @ 1 mL/min
    • 45 s @ 0.5 mL/min
  • RS232 6 digit output (XX.XXXX mL/min) to printer or PC

Hardware

  • Size: 5.4 x 3.0 x 1.8 in (138 x 76 x 45 mm)
  • Weight: 12 oz (320 g)
  • 4-digit LCD with status symbols
  • Compatible with all standard HPLC solvents

Included

  • UKAS (equivalent to NIST) traceable single-point calibration at 1.0 mL/min (with additional points on request) valid for 1 year
  • Universal mounting kit (bench, wall, stack, bottle top)
  • Carrying case with foam inserts, cleaning kit, RS232 cable
  • Mains power supply (US/EU/UK options)

Uses for the FlowCal 5000

  • IQ/OQ/PQ and quarterly calibration
  • Troubleshooting, e.g. detect leaking piston seals or check valves
  • Service multiple systems with one portable meter

Flow meter recalibration

Annual recalibration is highly recommended. The flow meter is shipped with a traceable calibration certificate valid for one year. Even if your flow meter is still operating within specifications at the end of 1 year, it will be cleaned and adjusted for improved accuracy.

Recalibration must include a calibration point at 1 mL/min. You can request any number of additional calibration points between 0.05 - 25 mL/min. Standard 3-point recalibration is at 0.5, 1, and 2 mL/min.

Typical turnaround: about 2 weeks. Rush service is usually possible for an additional fee.

Recalibration steps

  1. Contact Tovatech for pricing and instructions.
  2. Ship your flow meter to our Plano, TX facility.
  3. Flow meter is cleaned, adjusted, and calibrated.
  4. Flow meter and new calibration certificate are shipped back to you.

Included data

  • As-found data at 1 mL/min (additional points by request)
  • UKAS (equivalent to NIST) traceable certificates

Mini Thermal printer

FlowCal flow meter thermal printer

Printer Features

  • RS232 interface with rechargeable battery
  • Prints readings to four decimal places
  • Compact, easy paper loading, rugged case
  • Powered from the flow meter supply; no extra PSU needed

HPLC Flow Meter FAQs

How accurate and precise is the FlowCal 5000?

It is accurate to ±1% of reading over the linear range of 0.05-25 mL/min, with a standard deviation under ±0.5%.

What is the flow meter resolution?

The flow meter displays 4 digits (e.g. 0.050 mL/min), but when using an RS232 interface, the readings are output to 4 decimal places (e.g. 0.0500 mL/min).

How often should I recalibrate?

Annually. Flow meters are shipped with traceable certificates valid for 1 year. Call to schedule and get shipping instructions.

Can it run inline under pressure?

No, the flow meter is open to atmospheric pressure and cannot be used inline under pressure. It is connected to your HPLC system at the end of the line where the solvent exits at atmospheric pressure.

Is it compatible with my HPLC?

Yes, it is compatible with any HPLC system operating with a standard HPLC solvent at a flow rate between 0.05 and 25 mL/min. If you have unusual solvents, call so we can confirm.

How do I connect the flow meter to my HPLC system?

The flow inlet is a 1/16 inch diameter PEEK tube and can be connected with a standard fitting for HPLC tubing. Connecting the flow meter is easy.

What is the recalibration turnaround?

Typically about 2 weeks. Rush service is usually possible for a fee; call to arrange.

Does recalibration include as-found data?

Recalibration includes as-found data only at 1 mL/min. As-found data at other flow rates can be provided at additional cost, but it must be included on your recalibration purchase order.

Do I need the thermal printer?

It is optional. If you need hardcopy audit trails or field printing, we recommend it.

Product Brochure

Detailed product overview with features, benefits, and application examples.

Download Product Brochure

Need help with HPLC flow meter selection or recalibration?

Tell us your flow range, solvent compatibility, and calibration interval. We'll help you select the right meter and schedule traceable recalibration.